In addition to the operations above, you can specify the following settings on the Event To

tabbed page.

• Switch between absolute and relative time (Time Axis menu)

• Switch the time display format (Time Axis menu)

• Show or hide data numbers (Time Axis menu)

• Search for alarms (Edit menu or search bar)

• Clear cursors (Edit menu)

• Select all data (Edit menu)

• Add marks (Edit menu or toolbar)

• Delete marks (Edit menu or toolbar)

• Reset marks (Edit menu)

• Search for marks (Edit menu or search bar)

Explanation

The items on the Event To tabbed page and their descriptions are provided below.

Name

Display

Event Type

Tag Number

Available?

Loop Type

Description

TIME EVENT

Time event

TIME EVENT

[ TagNo ]

PV event

Yes

Tag number string

INT- [ LoopNo ]

No

Internal

INT-[loop number]

EXT- [ LoopNo ]

External

EXT-[loop number]

Loop numbers are displayed in two digits. A zero is inserted in the tens digit for numbers

less than 10.

Level

Displays L01 to L16.

Type

Display

Description

OFF

Event off or event setting release

TM

Time event

PVH

Measurement high limit

PVL

Measurement low limit

DVH

Deviation high limit

DVL

Deviation low limit

DVO

Deviation out limit

DVI

Deviation in limit

SPH

Setting high limit

SPL

Setting low limit

OTH

Output high limit

OTL

Output low limit

???

Displayed in cases other than the above

Event ON and Event OFF

Displays the time an event occurred and the time the event was released in the specified

format.