Installation, Location – Yokogawa Button Operated MV2000 User Manual

Page 20

Advertising
background image

Installation

Location

Install the MV indoors in an environment that meets the following conditions:

• Temperature of 0 to 40°C

Install the MV in a location where the temperature is 0 to 40°C, and the humidity is 20

to 80% RH (5 to 40°C). Only use the MV when there is no condensation on it.

Note

Condensation may form when moving the MV from a low temperature/humidity environment to
a high temperature/humidity environment, or when there is a sudden change in temperature.
Temperature/humidity changes may also result in thermocouple measurement errors. In these
kinds of circumstances, wait for at least an hour before using the MV, to acclimate it to the
surrounding environment.

• Altitude of 2,000 m or below

• Good ventilation

To prevent overheating, install the MV in a well-ventilated area. We recommend that

you leave 50 mm or more of space around the top, left, and right of the MV.

• Not much mechanical vibration

Install the MV in a location without much mechanical vibration. Placing the MV in a

place that is subject to large levels of mechanical vibration will not only put added
stress on its components, it may also impede ordinary measurement.

• Flat surface

Install the MV on a flat surface, neither leaning to the left nor to the right. The MV can

be tilted if the stand is used.

Do not install the MV in the following places:

• Outdoors

• In an environment with flammable or explosive gases, steam, or dust

(dangerous places)

• In direct sunlight or near heating devices

Install the MV in a place that is near room temperature (23°C) and that is not subject

to very much temperature fluctuation . Placing the MV in direct sunlight or near
heating devices can cause adverse effects on the internal circuitry.

• In an environment with excessive amounts of soot, steam, moisture,

dust, or corrosive gases

Soot, steam, moisture, dust, and corrosive gases will adversely affect the MV and

should be avoided.

• Near strong magnetic fields

Install the MV in a place with a magnetic field of less than 400 A/m. Do not bring

magnets or instruments that produce electromagnetic fields close to the MV. Operating
the MV near strong magnetic fields can cause measurement errors.

• Where the display is difficult to see

The MV uses an LCD screen so viewing of the display from an extreme angle is

difficult. Install the MV so that the user can view the display directly from the front.

20

IM MV1000-02E

Advertising
This manual is related to the following products: