3 controls and installation, 1 electrical installation, 1 pin assignment of the connector – LumaSense Technologies ISR 6 Advanced User Manual

Page 15

Advertising
background image

ISR 6 Advanced Manual

Controls and Installation · 15

3 Controls and Installation

3.1 Electrical Installation

The pyrometer is powered by 24 V DC ± 25% (very well stabilized, ripple max. 50 mV). It is
important to ensure correct polarity when connecting the device to the power supply.
To meet the electromagnetic requirements (EMV), a shielded connecting cable must be used.
LumaSense offers connecting cables, which are not part of the standard scope of delivery.
The shield of the connecting cable has to be connected only on the pyrometer’s side. If the
connecting cable is extended, the shield of the extension also needs to be extended. The shield
must be open on the power supply side (switch board), to avoid ground loops.
The connecting cable has wires for the power supply, interface, analog output, external laser
switch, and external clear of maximum value storage via contact and 12 pin connector. The cable
includes a short adapter cable with a 9-pin D-SUB connector. This connector may be used in
combination with the RS485 to USB adapter.
Once the instrument has been connected to the power supply, it is immediately ready for use.
Although it does not need to be warmed up, it does need to run for approximately 15 to 30

minutes before achieving full accuracy. The instrument can be switched off by interrupting the
power supply or unplugging the electrical connector.

Attention: When connecting the power supply, ensure the polarity is correct.

3.1.1 Pin assignment of the connector

Pin Color Function

K

white

+24 V DC power supply

A

brown

0 V DC power supply

L

green

+ I

out.

analog output

B

yellow

– I

out.

analog output

H

gray

Targeting light activate /
deactivate via external switch
(bridged with K)

J

pink

External clearing of max. value
storage (bridge to K), hold
function, or output for “dirty

window” monitoring (*see
notes below)

G

red

DGND (GND for interface)

F

black

B1 (RS485) or RxD (RS232)

C

violet

A1 (RS485) or TxD (RS232)

D

gray/pink B2 (RS485) (bridged with F)

E

red/blue A2 (RS485) (bridged with C)

M

orange

Screen only for cable extension,
don’t connect to the
switchboard