9 troubleshooting – LumaSense Technologies ISR 6 Advanced User Manual

Page 45

Advertising
background image

ISR 6 Advanced Manual

Troubleshooting · 45

9 Troubleshooting

Symptom

Probable Cause

Comments

No analog output even
if the display shows a
temperature above
lower range limit

Reversed leads to ISR 6.

The ISR 6 is diode protected if
power leads are reversed.
See section 3.1

Reversed leads to other
instruments
in the current loop.

Open circuit in lines connecting
all instruments in current loop.

Total burden higher than 500

Ohms

Insufficient supply voltage (See
wiring diagram).

Temperature reading
differs between built-in
display and external
instrumentation

Pyrometer output setting differs
from external instrument’s input
setting

Check the 0 to 20 mA or
4 to 20 mA on both sides
Check the sub range setting on
pyrometer against the range

setting on the external
instrument

Temperature readings
are too low

Target size too small.
Target is not at correct focus
distance.

See sections 3.4, 3.5, and 5.2.11
With small target/close focus
models, aiming may require X-Y
positioning table for fine

adjustment during aiming.

Incorrect alignment of the
pyrometer to the object.

Realign pyrometer in 1-channel
mode (mono) to achieve the
maximum temperature signal.

Dirty window or lens.

See sections 4.9 and 6.1

Target temperature too low.

Target emissivity much lower

than expected.

Even 2-Color instruments

depend on actual target
emissivity to supply sufficient
energy to the sensor.
See sections 4.2, 4.3, 4.4, 4.8,
and 6.1

Incorrect slope adjustments.

See sections 6.3 and 5.2.4

Faulty ISR 6

Contact LumaSense. See section
2.5