LumaSense Technologies ISR 6 Advanced User Manual

Page 47

Advertising
background image

ISR 6 Advanced Manual

Troubleshooting · 47

Symptom

Probable Cause

Comments

Noisy readings (fast
fluctuations).

Target moves more than 95%
out of field of view. Conveyor
type applications will also cause
fluctuations as targets move by

the field of view.

Application may require
Maximum to be stored. Apply
t

clear

settings. See section 4.6.2.

Target may be changing
temperature quickly.
(Thermocouples do not respond
as fast as the ISR 6).

Speed of response may be
slowed down

Apply t

90

settings. See section

4.5.

Target may be reflecting nearby
objects that have fluctuating
temperatures.

Loose connection in wiring.

Electrical noise lead wires caused
by intense magnetic fields,
improper selection of inter-
connecting cable.

Noisy readings (fast

fluctuations).

Cable run into in same conduit as

AC power lines.

Improper grounding of cable
shield and/or ISR 6 housing.

Flame or reflections of flame
may be entering the field of
view.

Extremely dense steam, smoke,
thick dust, personnel or moving

machinery intermittently
blocking 95% of optical path of
ISR 6.

Non-uniform emissivity of large
moving targets.

Slower response speed is
required.

Faulty ISR 6.

Contact LumaSense. See section
2.5

Laser Targeting Light
fails

The instruments internal
temperature is higher than 55°C.

Use cooling jacket.
See section 2.6.2