ISR 6 Advanced Manual

Settings / parameter descriptions · 27

you to program all pyrometers at the same time, regardless of the addresses that have already
been assigned. If the address of a pyrometer is unknown, it is possible to communicate with it
using the global Address 99 (connect only one pyrometer).

4.14 Focused Distance

The focused distance value can be adjusted at all distances

between 210 mm and 5000 mm using the Focus Adjustment
Screw on the back of the device.
The LED Distance Indicator Light will turn red and the
focused measuring distance in mm will automatically be
shown on the Digital Display within a few seconds of
making an adjustment using the Focus Adjustment Set Screw.
The focused distance can be viewed at any time through the InfraWin software or by using the
UPP Data Format commands.

4.15 Baud Rate

The transmission rate of the serial interface in Baud (Bd) is dependent on the length of the

cable. A maximum cable length for 19200 Bd with RS485 is 2 km. The baud rate is reduced by
50% if the transmission distance is doubled.

Note: Settings for Baud Rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, or
115200.

4.16 Pyrometer Internal Temperature

The internal temperature of the pyrometer can be read through the PC interface using the
InfraWin software or by using the UPP Data format commands. It is a few degrees higher than
the ambient temperature due to the heat generated by the electronics.
When using the Laser Targeting Light, the targeting light is switched off automatically if the
internal temperature of the pyrometer exceeds 55 °C. This safety feature is used to prevent
damage to the laser. It can only be used again once the temperature falls below 55 °C.