Correct weighing – Seca 665 User Manual

Page 22

Advertising
background image

22

Checking that the contents of the calibration counter are correct

This seca scale is calibrated. Calibration may be performed only by authorised centres.

To ensure this, the scale is fitted with a calibration counter which records any change to

data relevant to calibration. If you want to check whether the scale is properly calibrated,

proceed as follows:
– If necessary, switch off the scale.
– Keep any key depressed and start the

scale.
The current contents of the calibration

counter are shown on the display for a

few seconds.

– Compare the output contents of the

calibration counter with the number

shown on the calibration counter stick-
er.
For valid calibration, both numbers must

match. If sticker and calibration counter

do not match, then re-calibration is re-
quired. Contact your service agent or

seca Customer Service.

If a supplementary calibration is necessary,

then the supplementary calibration mark il-
lustrated opposite is used to record the

calibration counter position instead of the

seca calibration mark shown above. This

mark is secured with an additional seal by

the person authorised to perform the sup-
plementary calibration. The supplementary

calibration mark is available from seca

Customer Service under the number

14-05-01-886.

Correct weighing

– Press the Start key with no load on the

scale.

se(a

,

8.8.8.8.8.

and

0.0

appear consecu-

tively in the display. The scale is then auto-
matically set to zero and ready to use
– Switch weighing range if necessary

(see page 23).

– Move your chair onto the scale and

keep still.
Note

The maximum load on each individual

corner of the weighing platform is

200 kg.

– Read the weighing result off the digital

display.

Both numbers must match.

Advertising