7. Cleaning

Clean the rubber surface and the housing as required using a domestic cleaning agent or

commercially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like for

cleaning. Such substances can damage the high-quality surfaces.

8. What do I do if…

… no weight display comes on?
– Is the scale switched on?
– Check the mains connection and the connection of the battery pack
– Is the plug contact on the scale correctly connected?
– Is the plug contact for connecting the display to the scale correctly connected?

0.0

does not appear before the weighing operation?

– Press the Start key again – there must not be any load on the scale – and only its feet

should be in contact with the floor.

… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.

… the display shows

?

– Battery voltage is dropping. Charge the rechargeable batteries in the next few days.

batt

appears in the display?

– The rechargeable battery is so severely discharged that weighing is no longer possible.

Recharge the battery.

St0p

appears in the display?

– Maximum load has been exceeded.

… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-

tected a steady weight value. The display will then stop flashing.

– If there has been no activation of a function beforehand, remove the load from the scale

and wait until

0.0

is displayed, then weigh again.

… the display

appears?

– The ambient temperature of the scale is too high or too low. Place the scale in an

ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the scale

to adapt to the ambient temperature and then weigh again.

… the weighing range won’t switch?
– Take the load off the scale if necessary.
– Ensure that no special function is active.

… the display

E

and a number appear?

– Switch off the scale with the Start key and start the scale again. The scale will then work

normally again.

If this is not the case, interrupt the power supply by briefly removing the rechargeable

battery. If this measure is equally unsuccessful, inform the Service department.