Setting damping

Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display re-
acts. You can set damping to suit requirements.
– Press the

FUNC key and keep it de-

pressed for 2 seconds.
FIL and the current setting appear in

the display:
0 = slight damping

➞ fast determination of weight

1 = average damping

➞ normal determination of weight

2 = strong damping

➞ slow determination of weight

– Press an arrow key if you want to

change the set value.

– Press the

FUNC key again to store the

displayed value. The scale is then back

in normal weighing mode.

The settings are retained even once

the scale is switched off.

6. Storing the scale

The scale can be stored upright to save space.
– Undo the knurled nut on the hinge of

the column.

– Pull out the knurled screw and fold over

the column.

– Re-engage the knurled screw.
– You can now prop the scale up on the

column.

– The scale can now be moved.

The scale can be placed on the bottom rail

for short-term storage. Secure the scale

so that it does not fall over when stored in