Maintenance, Technical data – Seca 665 User Manual

Page 28

Advertising
background image

28

9. Maintenance

Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the cali-
bration counter does not match the number on the applicable calibration counter
mark (cf. page 22).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.

10. Technical data

Dimensions: (width, length, height)

Weighing plattform:

800 x 965 x 65 mm

Scale in upright position:

193 x 986 x 1102 mm

Weight

approx. 32.3 kg

Temperature range

+10 °C to +40 °C

Calibration

as per Directive 2009/23/EC

approved for medical use, Class III

Medical product as per Directive
93/42/EEC

Class I

Colour

silver grey/black

Maximum load

Weighing range 1

200 kg

Weighing range 2

300 kg

Minimum load

Weighing range 1

2 kg

Weighing range 2

4 kg

Graduations

Weighing range 1

100 g

Weighing range 2

200 g

Accuracy at first calibration

Weighing range 1:
to 50 kg:

± 50 g

50 kg to 200 kg:

± 100 g

Weighing range 2:
0 to 100 kg:

± 100 g

100 kg to 300 kg:

± 200 g

Power supply:

mains unit and built-in rechargeable
battery pack

Advertising