GW Instek AFG-100/200 Series User Manual User Manual
Page 60

AFG-125/225 Series Options Manual
60
2. From the bottom menu press ARB.
3. From the side menu turn ARB On.
Note
The function generator also will not allow ARB to be
turned on when sweep, burst or a modulation function
is active. Turn off any other active modes before using
ARB mode.
4. Press Edit from the side menu.
Point/Line
Editing
5. Press Edit Method and choose Point
Line.
6. Press Action to begin editing the
waveform using the Point/Line
method.
7. Use the side menu to create a point or vertical
line:
Press Address to set the starting
address of the line or point.
The maximum address depends on
the Length setting below:
Max address = 4096 – Length
Range:
0 ~ 4095
Press Length to set the length of the
line. The maximum length depends
on the address set above:
Max length = 4096 – Address
Length