Setting arb frequency, rate, amplitude & offset – GW Instek AFG-100/200 Series User Manual User Manual
Page 68
Advertising

AFG-125/225 Series Options Manual
68
Note
* The maximum length depends on the start address
setting.
The maximum length = 4096 – Start address.
3. Press Confirm to set the chosen
output section.
4. See page 29 to set the output
settings and turn the output on.
Setting ARB Frequency, Rate, Amplitude & Offset
Background
The rate, frequency, amplitude and
(amplitude)offset output parameters can also be
set.
Steps
1. From the ARB menu press more 1
of 2.
2. Select the parameters from the side menu:
Sets the frequency* of the ARB
waveform.
Sets the amplitude** of the
waveform:
1mVpp to 2.5Vpp (into 50Ω)
2mVpp to 5Vpp (open-circuit)
Sets the Offset** of the ARB
waveform:
±1.25Vpk ac +dc (into 50Ω)
±2.5Vpk ac +dc (Open circuit)
Advertising