Rear panel overview – GW Instek LCR-8000G Series User Manual
Page 10

GETTING STARTED
19
19
19
19
Rear Panel Overview
GPIB
RS232
WARNING
SER.NO. LABEL
AC 115 / 230V
50 / 60Hz
FUSE 250V T 3A
1
15
V
GPIB
Port
RS-232C
Port
Display
Contrast
Knob
Voltage
Selector
Fuse
Holder
Mains
Socket
GPIB port /
RS-232C port
GPIB
RS232
Accepts remote control cables.
GPIB: 24-pin female
RS-232C: DB-9 pin male
See page107 for remote control
details.
Display contrast
knob
Sets the display contrast level. See
page22 for details.
LCR-8000G Series User Manual
20
20
20
20
Voltage selector /
Fuse holder /
Mains socket
11
5V
The voltage selector sets the AC
mains voltage:
AC 115V (+10% / -25%), AC 230V
(+15% / -14%) (Selectable),
50/60Hz.
The fuse holder contains the main
fuse, T3A/250V. For fuse
replacement details, see page130.
The mains socket accepts power
cable. See page22 for power-up
details.