GW Instek LCR-8000G Series User Manual
Page 36

PASS-FAIL MODE
71
71
71
71
Parameters
No. of step
Maximum 30 for each program
No. of program Maximum 64
Drive Voltage
(1mV step)
10mV ~ 2V (DC or AC≤3 MHz)
10mV ~ 1V (AC>3 MHz)
Frequency
20Hz ~ 1MHz (LCR-8101G)
20Hz ~ 5MHz (LCR-8105G)
20Hz ~ 10MHz (LCR-8110G)
Bias
Reserved item: internal use only
Speed
Max, Fast, Med, Slow
Hi / Lo Limit
Follows the measurement range
Delay
0 ~ 9999ms, 1ms step
Single trigger
Program runs when the Trigger
key or F1 (Start) is pressed.
Automatic
trigger
Program runs when the LCR-
8000G detects the DUT.
LCR-8000G Series User Manual
72
72
72
72
Configure beep setting
Background
The beep sounds when the pass/fail test result
matches the setting: failed or passed.
Panel operation
1. Press the Menu key, then F5
(System). The system configuration
appears.
Menu
F 5
2. Press the Up/Down key and move
the cursor to Beep.
Beep
:
OFF
3. Press the Left/Right key to select the
beep setting, Off, Pass, or Fail.
Off
Beep is turned Off
Pass
Beeps when the test result is pass
Fail
Beeps when the test result is fail