GW Instek GPI-700A Series Programming Manual User Manual
Page 14

GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 25 ⎯
*WAI
(no query form)
Function:
WAI prevents the programming instrument from executing further
commands or queries until all pending operations are finished.
Syntax:
*WAI
:GROup
Function:
Set or query the value of group.
Syntax:
:GROup <NR1>
:GROup?
<NR1>is in the range from 0 through 10.
* 10: COM mode
Return:
<NR1>
Examples:
:GROup 3 set the group 3.
:GROup? return 3 if the group setting is 3.
:STEP
Function:
Set or query the value of step
Syntax:
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 26 ⎯
: STEP <NR1>
:SETP?
<NR1> is in the range from 1 through 16.
*0: COM mode
Return:
<NR1>
Examples:
:SETP3 set the step 3
:SETP? return 3 if the STEP setting is 3.
:FUNCtion:TEST:MODe
Function:
Set or query the test function ( 0:EMPTY 1: ACW 2: DCW 3: IR 4: Cnt (GB for 745A) 5: CAC
6: CDC 7: HOLD ). Only Group 0 has CAC and CDC function, but Group 0 does
not have HOLD function.
Syntax:
:FUNCtion:TEST:MODe <NR1>
:FUNCtion:TEST:MODe?
<NR1> is in the range from 0 through 7.
Return:
<NR1>
Examples:
:FUNCtion:TEST:MODe 3 set the test function to IR.