GW Instek GPI-700A Series Programming Manual User Manual
Page 16
Advertising

GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 29 ⎯
V/I:
Test Function
Unit (2 bytes)
ACW
KV
DCW
KV
IR
V
Cnt (GB)
A
CAC
KV
CDC
KV
HOLD
PAUSE
Meas:
Test Function
Unit (4 bytes)
ACW
mA
DCW
mA
IR
MOhm
Cnt (GB)
mOhm
CAC
mA
CDC
mA
HOLD
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 30 ⎯
Time:
Time (1 byte)
Value (5 bytes)
Unit (1 bytes)
Ramp
005.0
sec
Test
005.0
sec
Examples:
User the command :MEAS3? after measurement
Display for PC 10,1,0.100kV,01.00mA, T005.0S
10=Pass
1=ACW
0.100kV=Mesuremett voltage 100V
01.00mA=Mesurement current 1mA
T005.0S=Test time 5 sec
:UTILity:AC:FREQuency
Function:
Set or query the ACW CAC frequency ( 0:50Hz, 1:60Hz )
Syntax:
:UTILity: FREQuency <Boolean>
:UTILity: FREQuency?
50 (50Hz) or 60 (60Hz)
Return:
50 | 60