GW Instek GPI-700A Series Programming Manual User Manual
Page 23

GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 43 ⎯
:DCWStanding:CMINimum <NR2>
:DCWStanding:CMINimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:CMINimum 1.00 set the DCW test minimum current to 1.00mA.
:DCWStanding:CMINimum? return 01.00 if the DCW test minimum current
setting is 1.00mA.
:DCWstanding:CMAXimum
Function:
Set or query the DCW test maximum current
Syntax:
:DCWStanding:CMAXimum <NR2>
:DCWStanding:CMAXimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:CMAXimum 1.00 set the DCW test maximum current to
1.00mA.
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
⎯ 44 ⎯
:DCWStanding:CMAXimum? return 01.00 if the DCW test maximum current
setting is 1.00mA.
:DCWStanding:RTIMe
Function:
Set or query the DCW ramp time.
Syntax:
:DCWStamding:RTIMe <NR2>
:DCWStamding:RTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:RTIMe 5.0 set the DCW ramp time to 5 sec.
:DCWStanding:RTIMe? return 005.0 if the DCW ramp time setting is 5 sec.
:DCWStanding:TTIMe
Function:
Set or query the DCW test time.
Syntax:
:DCWStamding:TTIMe <NR2>
:DCWStamding:TTIMe?
<NR2> Please refer to the specification