GW Instek GDM-8261 User Manual User Manual
Page 189

REMOTE CONTROL
189
[SENSe:]TEMPerature:FRTD:RESolution
Sets the 4W RTD measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX)
Example: SENS:TEMP:FRTD:RES MAX
Sets the 4W RTD resolution to the maximum.
[SENSe:]TEMPerature:FRTD:RESolution?
Returns the 4W RTD measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]TEMPerature:RTD:RESolution
Sets the 2W RTD measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX)
Example: SENS:TEMP:RTD:RES MAX
Sets the 2W RTD resolution to the maximum.
[SENSe:]TEMPerature:RTD:RESolution?
Returns the 2W RTD measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]VOLTage:DC:NPLCycles
Sets the integration time for DC Voltage measurements in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:VOLT:DC:NPLC 12
Sets the integration time to 12 PLCs for DC Voltage.
[SENSe:]VOLTage:DC:NPLCycles?
Returns the integration time for DC Voltage measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]CURRent:DC:NPLCycles
Sets the integration time for DC Current measurements in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:CURR:DC:NPLC 2
Sets the integration time to 2 PLCs for DC Current.