Calculate commands – GW Instek GDM-8261 User Manual User Manual

Page 190

Advertising
background image

GDM-8261 User Manual

190

[SENSe:]CURRent:DC:NPLCycles?

Returns the integration time for DC Current measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12

[SENSe:]RESistance:NPLCycles

Sets the integration time for 2W resistance measurements in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:RES:NPLC MIN
Sets the integration time to 0.025 PLCs for 2W resistance measurements.

[SENSe:]RESistance:NPLCycles?

Returns the integration time for 2W resistance measurements in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12

[SENSe:]FRESistance:NPLCycles

Sets the integration time for 4W resistance measurements in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:FRES:NPLC MAX
Sets the integration time to the maximum for 4W resistance measurements.

[SENSe:]FRESistance:NPLCycles?

Returns the integration time for 4W resistance measurements in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12

CALCulate Commands

CALCulate:FUNCtion

Sets the Advanced function.
Parameter: OFF | MIN | MAX | HOLD | REL | COMP | DB | DBM | STORE |
AVER | MXB | INV | REF
Example: CALC:FUNC REL
Sets the Advanced function to REL (relative)

CALCulate:FUNCtion?

Returns the current Advanced function.