Calculate commands – GW Instek GDM-8261 User Manual User Manual
Page 190

GDM-8261 User Manual
190
[SENSe:]CURRent:DC:NPLCycles?
Returns the integration time for DC Current measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]RESistance:NPLCycles
Sets the integration time for 2W resistance measurements in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:RES:NPLC MIN
Sets the integration time to 0.025 PLCs for 2W resistance measurements.
[SENSe:]RESistance:NPLCycles?
Returns the integration time for 2W resistance measurements in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
[SENSe:]FRESistance:NPLCycles
Sets the integration time for 4W resistance measurements in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.025, 0.1, 0.25, 1, 2, 12).
Parameter: NPLCycles(<NRf> | MIN | MAX)
Example: SENS:FRES:NPLC MAX
Sets the integration time to the maximum for 4W resistance measurements.
[SENSe:]FRESistance:NPLCycles?
Returns the integration time for 4W resistance measurements in PLCs
(power line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.025, 0.1, 0.25, 1, 2, 12
CALCulate Commands
CALCulate:FUNCtion
Sets the Advanced function.
Parameter: OFF | MIN | MAX | HOLD | REL | COMP | DB | DBM | STORE |
AVER | MXB | INV | REF
Example: CALC:FUNC REL
Sets the Advanced function to REL (relative)
CALCulate:FUNCtion?
Returns the current Advanced function.