Store measurement record – GW Instek GDM-8261 User Manual User Manual
Page 96

GDM-8261 User Manual
96
Store Measurement Record
Background
The GDM-8261 can log up to 9999 measurement results
(counts) which can be stored and recalled later for
analysis. Basic measurement statistics such as Maximum,
Minimum, Average value as well as Standard Deviation
are also recorded with the data.
Note: Previously recorded measurements will be
erased every time the store function is used or if
power is reset.
Data count
2 ~ 9999
Not applicable to
Store/recall measurement history is not applicable to the
Diode/Continuity tests
/
.
Store step
1. Press the Shift key followed by
the DCI (Store) key. The store
menu appears.
/ EXIT
SHIFT
DCI
STORE
2. Move the cursor using the
Left/Right key. Change the data
count using the Up/Down key.
HOLD
TRIG
3. Press the Enter key to confirm
editing and to go back to the
previous display.
AUTO
ENTER
V
m
S
DC
STO
STO
Indicates the measurement history is
stored