*ESR?
Returns and clears SESR (Standard Event Status Register).
Example: *ESR?
>198
Returns 198. SESR=11000110
*IDN?
Returns the manufacturer, model No., serial number and system version
number.
Example: *IDN?
>GWInstek,GDM8261,00000000,1.0
*OPC?
“1” is placed in the output queue when all the pending operations are
completed.
*OPC
Sets operation complete bit (bit0) in SERS (Standard Event Status Register)
when all pending operations are completed.
*PSC?
Returns power On clear status.
Return parameter: <Boolean>(0 | 1) 0=clear, 1=don’t clear
*PSC
Clears power On status.
Parameter: <Boolean>(0|1) 0=clear, 1=don’t clear
*RST
Recalls default panel setup (reset the device).
*SRE?
Returns the SRER (Service Request Enable Register) contents.
*SRE
Sets SRER contents.
Parameter: <NR1>(0~255)
Example: *SRE 7
Sets the SRER to 00000111.
*STB?
Returns the SBR (Status Byte Register) contents.
Example:*STB?
>81
Returns the contents of the SBR as 01010001.
*TRG
Manually triggers the GDM-8261.