5.8. POSITION

“INRUSH

A ”: INRUSH CURRENT MEASUREMENT

WARNING

• The maximum measurable DC or AC+DC current is 1000A. Do not

measure currents exceeding the limits given in this manual. Exceeding
these limits could result in electrical shocks to the user and damage to the
instrument.

• We recommend holding the clamp respecting the safety area created by

the hand protection (see Fig. 3).

• Currents <3A are zeroed.

By positioning the selector switch to “InRush A ”, the screen to
the side will appear.
The display shows the current settings for recording inrush
currents (see par. 5.2.4).

Mod Zero Run

Mem

50Hz 01/10


10 A


Fix 15A

1/1

19/01-17:00:00

Fig. 14: inrush current measurement

Press key F1 (Mod) to open the drop-down menu shown on the
screen to the side. At each subsequent pressure of key F1, the
cursor will scroll through the available items, as follows:
50Hz: 50Hz inrush current measurement;

60Hz: 60Hz inrush current measurement;

400Hz: 400Hz inrush current measurement;

DC: DC inrush current measurement;

Mod

Zero Run

OK

50Hz 01/10


10 A


Fix 15A

1/1


19/01-17:00:00

Help: it displays the connection between instrument and

system;

Esc: it closes the drop-down menu.

Select the desired mode and press key F4 (OK) to confirm.

50Hz

60Hz
400Hz
DC
Help