SOLAR300 - SOLAR300N

EN - 71

The selection parameters of harmonic analysis require the preliminary selection of
voltages or currents inside General Parameter sub-level. The below error screen are
shown in these cases:

12/09/2006 – 16:55:10

RECORDING SETTINGS

- Harmonics

+

Voltages

+

Currents

Voltage Anom.: 230V 6%-10%

Inrush current: 30A 1s fix

Flicker

Error: no selected current

CLP/EXP

TYPICAL

SEL

MODIFY

Fig. 125: No selected currents error screen

To solve the error situation of above screen select the “Current” inside “General
Parameter” sub-level (see § 5.4.2.7).

12/09/2006 – 16:55:10

RECORDING SETTINGS

- Harmonics

+

Voltages

+

Currents

Voltage Anom.: 230V 6%-10%

Inrush current: 30A 1s fix

Flicker

Error: no selected voltage

CLP/EXP

PREDEF.

SEL

MODIFY

Fig. 126: No selected voltages error screen

To solve the error situation of above screen select the “Voltages” inside “General
Parameter” sub-level (see § 5.4.2.7).