The transmitter is configured from a standard IBM compatible PC by using

the IPRO software version 4.31 or later. The latest version of IPRO can

be downloaded from http://www.inor.com. When the transmitter is

set from a PC no calibration is needed.

When calibrating/configuring the IPAQ-21LX/-22LX ATEX

certified version of IPRO-X cable must be used.

Please observe that the 2 channels have separate configuration

outlets so put the IPRO-X plug in outlet 1 or 2 depending on which

channel to configure.

The transmitter is polarity protected for the power supply and will not

be damaged by connecting with the wrong polarity, but the output

will be 0 mA.

The maximum load in the output loop depends on the supply voltage,

see ”DATA shortform”

”Isolation input/output/PC” mentioned in the data sheet, indicates IS

galvanic isolation like an isolating barrier. Therefore ordinary power

supplies with Um < 250 VAC can be used with IPAQ-21LX/-22LX.

Sensor wire check: the transmitter uses a pulsating technique to monitor sensor

break or sensor shortened to avoid measuring errors. This pulsating signal may

interfere with some electronic calibrators. By selecting ”Sensor Break = None”

in the IPRO software and download the configuration into the transmitter the

pulsating signal is turned off.

GENERAL INFORMATION

INTRODUCTION

IPAQ-22LX is an Intrinsic Safe, 2-channel universal and intelligent

2-wire rail-mounted transmitter for temperature and other low-level

measurement Ex-applications. IPAQ-21LX = 1 channel.

IPAQ-21L/-22L are the non-Intrinsic Safe versions for use in safe areas.

Universal 1- & 2-channel

Programmable 2-wire

Transmitter

CONFIGURATION

Install the IPRO software from our software CD or from our web site

www.inor.com

.

The IPAQ-21LX/-22LX transmitters must be powered, during configuration

(see CONNECTIONS Fig 1). The IPAQ-21L/-22L must be powered if you use

configuration cable IPRO-X. If the 4-pole, so called ”off-line” cable is used

(only for IPAQ-21L/-22L), you usually don’t have to power the transmitter

during configuration. For computers with only USB-connection, use a USB/

RS232 adapter.

For keeping the safety of the internal circuits in the

IPAQ-21LX/-22LX, always use an ATEX -approved IPRO-X

configuration cable when making a configuration.

Before configuring an operating IPAQ-21LX/-22LX make sure the

input wiring is not connected to hazardous area.

When choosing differential temperature measurements with Pt100 be sure

to select ”sensor break” = none.

For IPAQ-21L/-22L only, the transmitter can be configured during operation.

The output is frozen while transmission from PC to IPAQ takes place. When

transfer is done the transmitter uses the new parameters.

mm/inches

!

ATEX

The user instruction must be read prior to adjustment and/or installation.

All information subject to change without notice.

This product should not be mixed with other kind of scrap, after usage.

It should be handled as an electronic/electric device.

www.inor.com, www.inor.se

MEASURE OF SUCCESS

MEASURE OF SUCCESS

INOR Process AB, PO Box 9125, SE-200 39 Malmö, Sweden,

Phone: +46-40-31 25 60, Fax: +46-40-31 25 70, E-mail: [email protected]

INOR Transmitter OY, Unikkotie 13, FI-01300 Vantaa, Finland,

Phone:+358-10-421 7900, Fax: +358-10-421 7901, E-mail: [email protected]

INOR Transmitter GmbH, Am Bocksborn 5,

D-63571 Gelnhausen, Germany,

Phone: +49-6051 14807, Fax: +49-6051 14806, E-mail: [email protected]

KROHNE Temperature Division INOR, 7 Dearborn Road,

Peabody, MA 01960, USA,

Phone: +1-978-826 6900, Fax: +1-978-535 3882, E-mail: [email protected]