2 specifications, Buoy specifications, Rdo titan probe and sensor specifications – In-Situ Aquaculture Buoy Manual (purchased after 28 May 2013) User Manual

Page 6: Specifications

Advertising
background image

6

Specifications

Buoy Specifications

Operating temperature

0° to 50° C (32° to 122° F)

Storage temperature

-5° to 60° C (23° to 140° F)

Dimensions (W x H x D)

61 x 91.4 x 81.3 cm (24 x 36 x 32 in.) (W x H x D)

Weight

18 kg (40 lbs)

Buoy materials

Polyethylene float with PVC mountings and fittings

Radio type

Frequency hopping; 100 mW

Transmitter

900 mHz spread spectrum; 9600 baud; 11.3 km (7 mi) maximum
range, line-of-sight (Achievable range is dependent on site
characteristics.)

Solar panel

10 W

Battery

12 V, 12 Ah, sealed lead acid

Brush motor

12 VDC

Brush enclosure

PVC

Warranty

1 year from date of shipment

RDO Titan Probe and Sensor Specifications

Optical RDO Titan Dissolved Oxygen Probe

Sensor Type

Optical (luminescent) dissolved oxygen sensor using RDO Classic Sensor Cap

RDO Titan Probe

Range: 0 to 50 mg/L concentration

Accuracy: ±0.1 mg/L, 0 to 8 mg/L

±0.2 mg/L, 8 to 20 mg/L

±10% of reading, 20 to 50 mg/L

Resolution: 0.01 mg/L
Display resolution: 0.10 mg/L

Response time: T90 < 45 sec; T95 < 60 sec @ 25°C

Storage conditions: -5° to 60° C (23° to 140° F)

RDO Classic Sensor Cap

Usage life: 1 year typical from the first instrument reading (15 months of total usage)

Shelf life: 36 months from date of manufacture

Storage conditions: 1° to 60° C (33° to 140° F), in factory container

Temperature sensor

Range: 0° to 50° C (32° to 122° F)

Accuracy: ±0.1° C typical

Resolution: 0.01° C

Transmitter/local display

Optional, sold separately

IP rating

IP-67 with cap off, IP-68 with cap installed

Compliance

Heavy industrial, IEC 61000-6-2:2005

Salinity range

0 to 42 PSU, fixed

Maximum power
consumption

50 mA at 12 VDC

Warranty

Probe: 3 years from date of shipment

Advertising