Baro merge input—fixed correction – In-Situ Rugged TROLL 100/200 Operators Manual User Manual

Page 36

Advertising
background image

800-446-7488

36

www.in-situ.com

Baro Merge Input—Fixed Correction

If you select the Apply a fixed correction option, a single correction is applied to all
values in the log.

To use this correction method you need barometric pressure values from a reliable
source. Choose a single value that represents the actual ambient barometric pressure
during the time period the log was recorded. You also need to know the file names of
the logs you want to correct.

1. From the Tools menu, select Win-Situ Baro Merge.

2. Select the Apply a fixed correction option.

3. Enter the barometric correction value and select units from the drop-down menu.

4. Click the right arrow button.

5. Select the log files to which the correction will be applied and click the check mark

button.

6. Compensated data files can be viewed or exported from the Data tab.