Rugged barotroll specifications – In-Situ Rugged TROLL 100/200 Operators Manual User Manual

Page 7

Advertising
background image

800-446-7488

7

www.in-situ.com

Rugged BaroTROLL Specifications

General

Rugged BaroTROLL

Temperature ranges

1

Operational: 0-50° C (32-122° F)
Storage: -40-80° C (-40-176° F)
Calibrated: 0-50° C (32-122° F)

Diameter

2.62 cm (1.03 in.)

Length

14.43 cm (5.68 in.)

Weight

170 g (0.37 lb)

Materials

Titanium body; Delrin nose cone, hanger, backend

Output options

USB or RS232 via docking station; Modbus/RS485 or SDI-12 via Rugged TROLL
200 Cable

Battery type & life

2

3.6 V lithium; 10 years or 2M readings

External power

8-36 VDC

Memory

Data records

3

Data logs

1.0 MB

65,000

2 logs

Fastest logging rate

1 per minute

Fastest output rate

Modbus & SDI-12: 1 per second

Log types

Linear

Sensor Type/Material

Piezoresistive; Ceramic

Range

7.0 to 30.0 psi; 0.5 to 2 bar

Accuracy (FS)

4

±0.1% full scale (FS) typical

±0.3% FS max.

Resolution

±0.01% FS or better

Units of measure

Pressure: psi, kPa, bar, mbar, mmHg, inHg

Temperature Sensor

Accuracy

±0.3° C

Resolution

0.01° C or better

Units of measure

Celsius or Fahrenheit

Warranty

2 years

Footnotes

1

Temperature range for non-freezing liquids

2

Typical battery life when used within the factory-calibrated temperature range.

3

1 data record = date/time plus 2 parameters logged (no wrapping) from device

within the factory-calibrated temperature range

4

Across factory-calibrated pressure range and temperature ranges

Specifications are subject to change without notice.
Delrin is a registered trademark of E.I. du Pont de Nemours and Company.