parameters were selected for the absolute/non-vented log in the corrected
BaroMerge file that uses the compensation table values. This is intended to show
how the adjustments were done in the BaroMerge file. If you do not want to show
these adjustments, clear this option.

l

When the second option, Show time in UTC is selected, the compensation table
time stamp displays in Coordinated Universal Time (UTC) time, formally known
as Greenwich Mean Time (GMT).

l

If the third option, Ignore daylight saving time (DST) is selected, the
compensation table time stamp format without the daylight savings time
adjustment will be shown.

3. Build a table that contains at least two barometric pressure values.

4. Click the right arrow and select the absolute (non-vented) log file or files you intend

to correct.

5. Click the check mark and the barometric compensation is applied.

6. Compensated data files can be viewed or exported from the Data tab.

BaroMerge Output

Your original log file is not changed. A new, corrected log file with the same name and
path is created. The original “.wsl” extension is replaced by “-Baro Merge.wsl”.

Disconnecting an Instrument from the Software

Click the plug icon

in the lower-right corner of the screen to disconnect the

instrument from the software.

Disconnect the instrument from the TROLL Com Communication Device. Attach a
desiccant pack if you are using a vented cable.