In-Situ Aqua TROLL 400 Operators Manual User Manual

Page 27

Advertising
background image

800-446-1500

27

www.in-situ.com

5. Completely submerge the RDO Sensor into the solution.

6. Click OK, to start the calibration.

7. When the screen indicates that the calibration has reached stability, click Accept to

complete the calibration, or click Cancel to return to the previous calibration.

8. You can save or print the calibration report.

9. Click OK to complete the calibration.

10. Once calibration is complete, remove the instrument from the calibration cup and

rinse both thoroughly with clean water.