Sensor Specifications

Level, Depth, Pressure Sensor Specifications

Accuracy

Typical ±0.1% FS @ 15° C; ±0.3% FS max. from 0 to 50° C

Range

76 m (250 ft); absolute (non-vented)

Resolution

±0.01% FS or better

Sensor Type

Fixed

Response
Time

Instantaneous in thermal equilibrium

Units of
Measure

Pressure: psi, kPa, bar, mbar, mmHg, inHg
Level: mm, cm, m, in, ft

Methodology

Piezoresistive; ceramic

Conductivity Sensor Specifications

Accuracy

Typical ±0.5% + 1 μS/cm; ±1% max.

Range

5 to 100,000 μS/cm

Resolution

0.1 μS/cm

Sensor Type

Fixed

Response
Time

Instantaneous in thermal equilibrium

Units of
Measure

Actual conductivity (μS/cm, mS/cm)
Specific conductivity (μS/cm, mS/cm)
Salinity (PSU)
Total dissolved solids (ppt, ppm)
Resistivity (Ohms-cm)
Density (g/cm3)

Methodology

Std. Methods 2510 EPA 120.1