Orp sensor specifications – In-Situ Aqua TROLL 400 Operators Manual User Manual

Page 8

Advertising
background image

800-446-1500

8

www.in-situ.com

Dissolved Oxygen RDO Classic (Optical Sensor) Specifications

Accuracy

±0.1 mg/L; ±0.2 mg/L; ±10% of reading

Range

0 to 8 mg/L; 8 to 20 mg/L; 20 to 50 mg/L; Full operating
range: 0 to 50 mg/L

Resolution

0.01 mg/L

Sensor Type

Fixed with replaceable RDO Classic Sensor Cap (life: 1
year typical)

Response
Time

T90: <45 sec. T95: <60 sec.

Units of
Measure

mg/L, % saturation, ppm

Methodology

EPA-approved In-Situ Methods 1002-8-2009 1003-8-2009
1004-8-2009

ORP Sensor Specifications

Accuracy

±5.0 mV

Range

±1400 mV

Resolution

0.1 mV

Sensor Type

Replaceable pH/ORP combo sensor

Response
Time

<15 sec.

Units of
Measure

mV

Methodology

Std. Methods 2580