9 controller requirements and connections, Wiring overview, Stripped-and-tinned cable – In-Situ Aqua TROLL 400 Operators Manual User Manual

Page 32: Power connections, Controller requirements and connections

Advertising
background image

800-446-1500

32

www.in-situ.com

Controller Requirements and Connections

The instrument can be connected to a controller or logger for communication via the
following protocols.

l

SDI-12

l

RS485 Modbus

l

RS232 Modbus (with converter)

Wiring Overview

Refer to the diagrams on the following pages. Trim and insulate unused wires. The
shielded wire should be wired to a chassis ground or earth ground. See "Safety" on
page 5.

Stripped-and-Tinned Cable

Signal

Color

Ground/Return

Black

External Power

Red

No Connection

Brown

RS485 (-)

Green

RS485 (+)

Blue

SDI-12

White

Power Connections

The Aqua TROLL 400 requires an external 8 to 36 VDC power source. The red wire
must be connected to the positive terminal of the power source. The black wire must be
connected to the negative terminal of the power source, which is often referred to as the
system ground or return.