3. Place a water-saturated sponge in the bottom of the calibration cup. Place the
instrument into the calibration cup, and tap Start.
The calibration cup must be vented to barometric pressure. Make
sure the vented cap is installed.
4. When the calibration is stable, tap the Accept button.