8
Barometric Pressure Sensor Specifications (Battery Pack)
Accuracy
±3 mbar max.
Range
300 to 1100 mbar
Resolution
0.01 mbar
Sensor Type
Fixed
Response
Time
Instantaneous in thermal equilibrium
Units of
Measure
psi, kPa, bar, mbar, mmHg, inHg, Torr, atm
Methodology
Piezoresistive pressure sensor
Air Temperature Sensor Specifications (Battery Pack)
Accuracy
±2° C
Range
-20 to 70° C (-4 to 158° F)
Resolution
0.1° C
Sensor Type
Fixed
Response
Time
<30 sec.
Units of
Measure
Celsius, Fahrenheit
Methodology
EPA 170.1