3.2.2 Temperature Coefficient set up

The display will show:

The secondary display will show COEFF and will then scroll TEMPERATURE COEFFICIENT after
10 seconds.

Symbol will flash while adjustment is being made.

The first digit of the displayed value will be flashing. Adjustment of the flashing digit can be made
using the Up/Down arrow keys.

The Left/Right arrow keys are used to select the next adjusting digit which will flash when selected.

Once the full reading has been adjusted to the required value, press the STO key. The symbol will
disappear and the secondary display will momentarily show SAVED. The instrument display will
update and then show the Reference Temperature set up menu.

3.2.3 Reference Temperature selection

The display will show:

The secondary display will show REF T and will then scroll REFERENCE TEMPERATURE after 10
seconds. Select the reference temperature from the options available (18, 20 or 25°C) by scrolling
through the list using the Up/Down arrows.

Symbol will flash while adjustment is being made.

The Left/Right arrow keys and the CAL/CLR key have no function during this set up.

To accept the correct reference temperature value, press the STO key. The symbol will disappear
and the secondary display will momentarily show SAVED. The instrument display will update and
then show the TDS EC ratio set up menu.