Jenway 4510 User Manual

Page 19

Advertising
background image

16

Interfacing (continued)

The RS232 communications parameters on the computer or printer need to be set to match those of
the Model 4510, as detailed below:

1200 Baud

9600 Baud

7 Data Bits

8 data bits

Odd Parity

OR

No parity

1 Stop Bit

1 stop bit

Setting of these options is detailed in Section 3.6.6.

The Model 4510 supports both hardware (CTS/RTS) flow control and software XON/XOFF flow control.

Pressing the PRINT key outputs from the RS232 interface.

Sending an ASCII “D” to the 4510 causes a printout of the current displayed reading plus sample
number.

Sending an ASCII “C” causes a printout of the last calibration parameters.

Sending an ASCII "P" causes a printout of the stored readings.

6.3 Keypad Emulation

Keypad remote control using RS232 interface:

7

- Instrument On / Standby / Escape

1

- Calibrate / Memory Clear

9

- Print

8

- Up Arrow

2

- Down Arrow

4

- Left Arrow

6

- Right Arrow

3 or 5

- Enter / Store

451 050/REV A/03-03