Section 7

Troubleshooting

7.1Troubleshooting

Fault

Possible Cause

Action

No display

Check power supply

Check that correct 9V ac power supply
is connected and switched on.

Erratic display

Check power supply

Unit must be used with supplied 9V
acpower supply. Usage of other units will
cause the 4510 not to operate.

Unstable display

Conductivity cell defective

Replace conductivity cell.

Display permanently

Intermittent or no connection

Check cell connection to 4510.

under or over range

Conductivity cell defective

Replace conductivity cell

Contaminated solutions

Replace solutions

Intermittent display

Conductivity cell not fitted

Check connections.

correctly

Reading drifts

Conductivity cell stored dry

Allow to soak for 2 hours

CO

2

absorbtion by sample

Noticeable for low conductivity - do not
allow sample to stand in unstoppered
bottles.

Non linear readings

Conductivity cell not zeroed

Zero conductivity cell.

Poor reproducibility

Carryover between solutions

Rinse cell in distilled water between
measurements.

Will not print

Connection

4510 requires the RS232 cable to
connect the printer.

Paper out

The feed light on the printer will flash if
the unit requires paper.

Battery flat

Connect ac power supply.