For installations where the provided pre-dressed remote signal cable
is not used, both ends of the cable being used must be prepared per
the instructions in Tables 4 and 5 of document. The cable must be
connected at both ends per instructions on Tables 6 and 7 on Pages

14

17 and 18. Terminate wires at J-Box following Figure 9 on Page 16.
At housing end, wires should be terminated at 4-position terminal
block on rear of electronic module as shown in Figure 9.

1. Slide the knurled nut and then the rubber bushing onto

outer jacket of cable as shown at right. Next, remove outer
polyethylene jacket of cable to dimension shown.

2. Cut and remove braided copper shield to dimension shown

at right. This will expose the barrier (plastic) tape and foil
mylar that encloses the conductors.

3. Cut and remove the barrier tape, foil mylar and fillers to

dimension shown at right. This will expose two twisted
pairs of conductors (brown-yellow, orange-red) and an
uninsulated drain wire. The barrier tape under the copper
braid prevents the drain wire from shorting to the copper
braid shield.

4. Trim the uninsulated drain wire to dimension shown at right.

To expose bare conductors for termination, cut and strip
ends of the two twisted pairs to dimension shown.

5. Fold drain wire back onto the copper braid as shown at

right. Label outer cable jacket „VFM 3100 End” to avoid
confusion during installatin. Cable is now ready for
installation.

VFM 3100 End (Junction Box End) Procedures

Knurled nut

Outer polyethylene
jacket

Outer polyethylene
jacket

Rubber bushing

191 mm (7.5 in)

Copper braid

Knurled nut

Rubber bushing

25 (1.0)

165 mm (6.5 in)

Copper braid

Barrier tape and
foil mylar

Outer polyethylene
jacket

Knurled nut

Rubber bushing

Copper braid

Barrier tape and
foil mylar

Knurled nut

Outer polyethylene
jacket

Label cable
jacket

Rubber bushing

Copper braid

Barrier tape and foil mylar

25 (1.0)

165 mm (6.5 in)

Two twisted pairs

Uninsulated drain wire

Outer polyethylene
jacket

Knurled nut

Rubber bushing

Copper braid
do not tape

6.4 mm (.25)
cut and strip
4 places

25 (1.0)

165 mm (6.5 in)

25 (1.0)

Two twisted pairs

Uninsulated drain wire

Bare conductor (4 places)

Two twisted pairs

Uninsulated drain wire

VFM 3100

end

Table 4. Preparation of Remote Signal Cable