4.6.2. Replacing the Sensor

NOTE: Before beginning the assembly procedure, verify that you
have the correct kit of parts.

The sensor kits consist of:

1 Sensor Assembly
1 O-ring
1 Gasket
1 Flow Dam
2 Tie Wraps

Field terminal
cover

Conduit
connection

Mechanical
connector
bolts (4)

Electrical housing

Cover

Cover slot (both covers)

Mechanical connector

O-Ring

Sensor assembly

Flow dam

Gasket

Body

Figure 28. VFM 3100 Assembly

The following steps apply to both Standard and Extended
Temperature Range VFM 3100:
•

If the flow dam has remained in the VFM 3100 body, remove it
before starting to re-assemble.

•

Slide O-ring over sensor lead and onto neck of sensor.
See Figure 29.

•

Carefully feed sensor lead through hole in mechanical connector

and gently pull sensor lead out of electrical housing until sensor

is touching the mechanical connector. See Figure 31.

•

Place the flat gasket over the sensor in contact with serrated
sealing surface. Center the gasket. Slide the flow dam into the
groove of the sensor.

•

Insert the sensor with the connector into the VFM 3100 body
and assemble the four bolts finger tight.

Advertising