Installation – KROHNE OPTISYS TUR 1050 EN User Manual
Page 17

INSTALLATION
3
17
OPTISYS TUR 1050
www.krohne.com
11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en
Mounting procedure
• Take a water-level, a pencil and ruler and use the dimensions in the following drawing to mark
the six mounting hole locations on the mounting surface.
Note: Do not use the drawing on the following page directly as mounting template as the
printed dimensions there differ from reality!
• Drill six holes into the mounting surface and insert six screw anchors into them.
• Use screws M4 / 3/16" to fix the junction box power assembly at first.
• Put the electronic device on top of the junction box power assembly and use screws M6 / 1/4"
to fix it.
INFORMATION!
The design of the device recommends that the junction box power assembly cradles in a hutch
below the electronic device. Therefore the correct procedure is to fix the junction box power
assembly first and then mount the electronic device on top of it.
Figure 3-2: Mounting filed terminal box