Installation – KROHNE OPTISYS TUR 1050 EN User Manual
Page 21

INSTALLATION
3
21
OPTISYS TUR 1050
www.krohne.com
11/2012 - 4002444301 MA OPTISYS TUR 1050 R02 en
3.4.6 Inserting and fixing the ultrasonic cuvette with flow-through holder
CAUTION!
Never insert a cuvette with visible moisture or water on the glass or the transducer into the
upper part of the device, because this could destroy or damage the electronics or the
transducer! Always clean and dry the cuvette with a soft cloth just before inserting it. The
dehumidification system cannot remove large droplets of water, only residual moisture!
Figure 3-5: Procedure for inserting and fixing the ultrasonic cuvette with flow-through holder
INFORMATION!
The device can only detect a new cuvette if it works in the normal operation mode ("AUTO"). If the
device operates correctly after inserting a new cuvette, "AUTO" will flash on the display.