2

PROFIBUS PA

6

H250 M40 ESK4

www.krohne.com

03/2014 - 4002251801 - AD H250/M40 PROFIBUS R01 en

Device calibrated for Mass Flow

The calibration units are used as default units.

Device calibrated for Level

The default unit for level is [%]. Calibration unit is used as default unit for board temperature.

2.4 Technical data

Slot

Slot

Slot

Slot

Module type

Module type

Module type

Module type

Measuring data

Measuring data

Measuring data

Measuring data

1

1

AI-FB

Volume flowrate

Mass flowrate (Default setting)

Board temperature

2

Totalizer-FB

Totalized volume flowrate

Totalized mass flowrate (Default setting)

3

Totalizer-FB

Totalized volume flowrate (Default setting)

Totalized mass flowrate

1 Selection of measuring data can be changed by modifying the channel parameter via acyclic PROFIBUS services.

Slot

Slot

Slot

Slot

Module type

Module type

Module type

Module type

Measuring data

Measuring data

Measuring data

Measuring data

1

1

AI-FB

Level (Default setting)

Board temperature

2

Totalizer-FB

2

Totalized volume flowrate (Default setting)

Totalized mass flowrate

3

Totalizer-FB

2

Totalized volume flowrate

Totalized mass flowrate (Default setting)

1 Selection of measuring data can be changed by modifying the channel parameter via acyclic PROFIBUS services.
2 Totalizing is not active if the device is calibrated for level measurement.

Hardware

Type

PROFIBUS MBP interface according to IEC 61158-2 with 31.25 kbit/s;
voltage mode (MBP = Manchester Coded Bus Powered)

Polarity

Independent of polarity at electrical connection

Base current

16 mA

FDE

Separate fault disconnection electronics provided
(FDE = Fault Disconnection Electronics)

Max. error current

23 mA

Start current after 10ms

Lower than base current

Electrical connections

Device power supply

Bus powered 9...32 VDC - not intrinsically safe
Bus powered 9...24 VDC - intrinsically safe

Basic current

16 mA

Maximum error current

23 mA

Start current after 10 ms lower than basic current
Polarity sensitivity

No

Minimum cycle time

250 ms

AD_H250_M40_PB_PA_R01_en_4002251801_PRT.book Page 6 Friday, March 21, 2014 1:14 PM