Installation, 8 instructions for oxygen applications – KROHNE OPTIBAR DP 7060 C EN User Manual
Page 22

3
INSTALLATION
22
OPTIBAR DP 7060 C
www.krohne.com
04/2014 - 4003292001 - MA OPTIBAR DP 7060 C R01 en
3.8 Instructions for oxygen applications
Oxygen and other gases can be explosive when brought into contact with oils, grease and
plastics, so the following measures must also be taken:
• All components of the plant, such as e.g. measuring devices must be cleaned according to the
requirements of BAM (DIN 19247).
• Depending on the seal material, certain temperatures and pressures must not be exceeded
in oxygen applications, refer to
Technical data on page 63.
CAUTION!
Devices for oxygen applications are sealed in PE foil and provided with a label "Oxygen! Use no
Oil". Remove this foil just before mounting the device! After removing the protection for the
process connection the label O
O
O
O
2222
will be visible on the process connection. Penetration of oil,
grease and dirt should be avoided. Danger of explosion!
.book Page 22 Friday, May 2, 2014 10:25 AM