Current output
In the current output menu, the saturation region for above or below a threshold is set. Under
"Current output min./max." these limit values can be set. The factory setting is 3.8 mA and 20.5 mA.
This corresponds to the NAMUR recommendation NE 43.
Mode
Output characteristic
0...100% = 4...20 mA or
0...100% = 20...4 mA
Additional bi-directional flow:
20...4...20 mA or
4...20...4 mA
Failure mode
≤ 3.6 mA, ≥ 21 mA, last valid measured value
Min. and
Max.
Min. current
3.8 mA, 4 mA
Max. current
20.5 mA, 20 mA
Lock
adjustment /
Unlock
adjustment
In this menu item, a 4-digit PIN can be activated, which protects against undesirable or unintended
changes of the settings. With a PIN active, remote access via software or other systems is also no
longer possible.
Run now