KROHNE BATCHFLUX 5015 K EN User Manual

Page 30

Advertising
background image

Installation and operating instructions

30 BATCHFLUX


5.6 Current output I


No HARDWARE!
Function must be set to “OFF”.

Fct. 1.5 CUR.OUTP. I (not available, no
HARDWARE !)
Press key

→ .

→ FUNCTION I = Select function for the
current output,
press key

→

• OFF (switched

off,

do not change setting,

no hardware available)

• 1 DIR.

(1 flow direction)

• 2 DIR.

(2 flow directions, F/R mode,

forward/reverse)

Select with keys

↑ and ↓.

Transfer to subfunction “RANGE I“ with key

↵. Exception: when “OFF“ selected, return to
Fct. 1.5 CUR.OUTP. I.

→ RANGE I = Select measuring range,
press key

→

• 0 - 20 mA
• 4 - 20 mA

Select with keys

↑ and ↓.

Transfer to subfunction “I ERROR“ with key

↵.

→ I ERROR = Set error value,
press key

→

• 0 mA
• 3.6 mA (only

possible

when

range “4 - 20 mA“

selected)

• 22 mA

Select with keys

↑ and ↓.

Press key

↵ to return to Fct. 1.5 CUR.OUTP.

I.

Please refer to Sect. 3.2
“Factory settings“.

Refer to Sect. 2.3 for connection diagrams,
and to Sect. 5.14 for characteristics.


5.7 Pulse output P


Fct. 1.6 PULS.OUTP. P
Press key

→ .

→ FUNCTION P = Select function for the
pulse output,
press key

→

• OFF

(switched off, no function)

• 1 DIR. (1

flow

direction)

• 2 DIR.

(2 flow directions, F/R mode,

forward/reverse)

Select with keys

↑ and ↓.

Transfer to subfunction “SELECT P“ with key

↵. Exception: when “OFF“ selected, return to
Fct. 1.6 PULS.OUTP. P.

→ SELECT P = Select pulse type,
press key

→

• 100-1000 Hz
• PULSE/VOL.

(pulses per unit volume,

flow)

• PULSE/TIME

(pulses per unit time

for 100 % flow)

Select with keys

↑ and ↓.

Transfer to subfunction “PULSEWIDTH“
with key

↵.

Note: when 100-1000 Hz selected,
return to Fct. 1.6 PULS.OUTP. P.

→ PULSEWIDTH = Select pulse width,
press key

→

• 50 mSec

F

max

= 10 Hz

• 100 mSec

F

max

= 5 Hz

• 200 mSec

F

max

= 2.5 Hz

• 500 mSec

F

max

= 1 Hz


for all

F

min

= 0.0056 Hz

(= 20 pulses / hr)


Select with keys

↑ and ↓.

Transfer to subfunction “VALUE P“ with key

↵

or return to Fct. 1.6 PULS.OUTP. P,
depending on selected pulse type in
subfunction “SELECT P“.

5

Advertising