KROHNE IFC 020 Converter EN User Manual

Page 40

Advertising
background image

Installation and operating instructions IFC 020

5/9

5.11 Primary head

Please refer to Sect. 3.2 “factory settings”

Fct. 3.02 FLOW METER

Press

→

key.

→

DIAMETER = set meter size (see instrument nameplate) press

→

key

Select size from table of meter sizes:

• DN 2.5 - 1000 mm equivalent to

1

/

10

- 40 inch

Select using

↑

or

↓

key.

Transfer to subfunction “FULL SCALE” with

↵

key.

→

FULL SCALE = set full-scale range, press

→

key.

Set as described in Sect. 5.1.

Transfer to subfunction “GK VALUE” with

↵

key.

Note: if “VALUE P” is displayed after pressing

↵

key.

PULSE/VOL. is set under Fct. 1.06 PULS.OUTP. P, subfunction “SELECT P”. Because the

full-scale range Q

100%

has been changed, the output frequency (F) of the pulse output is over- or

undershot:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

100%

Change pulse value accordingly, see Sect. 5.7 pulse output P, Fct. 1.06.

→

GK VALUE = set primary constant GK, press

→

key.

• 1.0000 - 9.9999 (note information on instrument nameplate, do not change setting !)

Change flashing digit (cursor) with

↑

or

↓

key.

Shift cursor 1 place to right or left with

→

or

←

key.

Transfer to subfunction “FIELD FREQ.” with

↵

key.

→

FIELD FREQ. = set magnetic field frequency, press

→

key

• 1/2 • 1/18 (1/2, 1/6 1/18 or 1/36 of power frequency,

• 1/6 • 1/36 see instrument nameplate, do not change setting)

Select using

↑

or

↓

key.

Transfer to subfunction “FLOW DIR.” with

↵

key.

(only for units with DC power supply, transfer to subfunction “LINE FREQ”).

→

LINE FREQ. = normal line frequency in your country, press

→

key.

(Please note, applicable for flowmeters equipped with DC in power supply!)

• 50 Hz

Select using the

↑

or

↓

key.

• 60 Hz

Transfer to subfunction “FLOW DIR.” with

↵

key.

→

FLOW DIR. = set flow direction, press

→

key.

• + DIR.

(for identification of flow direction, see “+” arrow on primary head;

• - DIR.

for F/R mode, identifies the “positive” flow direction)

Select using the

↑

or

↓

key.

Press

↵

key to return to Fct. 3.02 FLOW METER.

Zero check, see Fct. 3.03 and Sect. 7.1.

}