Installation and operating instructions IFC 020

7/1

7. Functional checks

7.1 Zero check Fct. 3.03

Switch off power supply before opening the housing
•

Set “zero” flow in the pipeline, but make sure that the measuring tube is completely

filled with fluid.

•

Switch on the system and wait 15 minutes.

•

Press the following keys for zero measurement:

Key

Display

Description

→

If “YES” set under Fct. 3.04 ENTRY CODE, key in

9-keystroke CODE 1 now:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

Fct. 1.00

OPERATION

2x

↑

Fct. 3.00

INSTALL.

→

Fct. 3.01

LANGUAGE

2x

↑

Fct. 3.03

ZERO SET

→

CALIB. NO

↑

CALIB. YES

↵

0.00

- - - - - / - - -

Flowrate displayed in set unit, see Fct. 1.04 DISPLAY,
subfunction “DISP. FLOW”.
Zero measurement in progress, duration approx. 15 - 90 seconds.
When flow “> 0” “WARNING” notice appears, confirm with

↵

key.

STORE NO

If new value not to be stored, press

↵

key (3x) 4x = return to

measuring mode.

↑

STORE YES

↵

Fct. 3.03

ZERO SET

Store new zero value.

(2x) 3x

↵

- - - - - - -

- - - - - / - - -

Measuring mode with new zero.

Key

Display

Description

→

If “YES” set under Fct. 3.04 ENTRY CODE, key in

9-keystroke CODE 1 now:

→ → → ↵ ↵ ↵ ↑ ↑ ↑

Fct. 1.00

OPERATION

↑

Fct. 2.00

TEST

→

Fct. 2.01

TEST Q

→

SURE NO

↑

SURE YES

↵

0

PERCENT

Current, pulse and status indication outputs indicate the

corresponding values.

± 10

PERCENT

↑

± 50

PERCENT

Select using

↑

or

↓

key

± 100

PERCENT

± 110

PERCENT

↵

Fct. 2.01

TEST Q

End of test, actual measured values again present at outputs.

(2x) 3x

↵

- - - - - - -

- - - - - / - - -

Measuring mode

7.2 Test of measuring range Q, Fct. 2.01

Switch off power supply before opening the housing !
•

For this test a measured value can be simulated in the range of -110 to +110 percent

of Q

100%

(full-scale range set, see Fct. 1.01 FULL SCALE).

•

Switch on the system.

•

Press the following keys for this test: