KROHNE IFC 020 Converter EN User Manual
Page 57
Installation and operating instructions IFC 020
If you need to return your flowmeter to Krohne,
please refer to last but one page of these Instructions !
Full-scale range
Q
100%
= 200 m
3
/hr (Fct. 1.01)
Meter size
DN
= 80 mm = 3“ (Fct. 3.02)
Current at Q
0%
I
0%
= 4 mA
Q
100%
I
100%
= 20 mA
}
(Fct. 1.05)
Pulses at Q
100%
P
100%
= 200 pulses/hr (Fct. 1.06)
Primary head constant
GK
= 3.571 (see nameplate)
Constant
(V
in m3)
(t
in hr)
K
= 7074 (see Table)
(DN in mm)
Calculation of “ X ” and position of “ Y ”
X =
Q
100%
x K
=
200 x 7074
= 61.905
GK x DN
2
3.571 x 80 x 80
Y = 40, position of switch Y, see front panel GS 8A
(comes closest to X value and is smaller than X).
Calculation of setpoint readings I and f
I = I
0%
+
Y
(I
100%
- I
0%
) = 4 mA +
40
(20mA - 4mA) = 14.3 mA
X 61.905
Deviations are permissible between 14.1 and 14.6 mA (equivalent to ± 1.5 %).
f =
Y
x P
100%
=
40
x 200 pulses / hr = 129.2 pulses/hr
X 61.905
Deviations are permissible between 127.3 and 131.1 pulses/hr (equivalent to ± 1.5 %).
7/8
Example 7.5.4