Operation – KROHNE OPTIWAVE 5200 FOUNDATION FIELDBUS User Manual

Page 33

Advertising
background image

OPERATION

4

33

OPTIWAVE 5200 C

www.krohne.com

09/2013 - 4002500101 - AD FF OPTIWAVE 5200 R01 en

DIAGNOSTIC_VALUE

Diagnostic Value

R

Current field diagnostic condition.

Uninitialized

Setting:

•

(00) Check Function

•

(01) Cond01

•

(02) Cond02

•

(03) Cond03

•

(04) Cond04

•

(05) Cond05

•

(06) Cond06

•

(07) Bad Measurement quality

•

(08) Signal Strong

•

(09) Signal Weak

•

(10) Empty Spectrum invalid

•

(11) Sensor No Signal

•

(12) Sensor Microwave Error

•

(13) Temperature Above +75°C

•

(14) Temperature Below -35°C

•

(15) Overfill Warning

•

(16) Level Lost warning

•

(17) Temperature out of range (warning)

•

(18) Local operation

•

(19) Overfill Error

•

(20) Level lost error

•

(21) Temperature out of range Error

•

(22) Converter Voltage Error

•

(23) Converter ROM Error

•

(24) Converter RAM Error

•

(25) Converter EEPROM Error

•

(26) Sensor not compatible

•

(27) Internal Communication Error

•

(28) Sensor Voltage Error

•

(29) Sensor ROM Error

•

(30) Sensor RAM Error

•

(31) Sensor EEPROM Error

ENABLE_DISABLE

Simulate En/Disable

RW

Enable/disable simulation.

Uninitialized

Setting:

Unitialized / Disabled / Active

FD_VER

FD_VER

FD_VER

FD_VER

Revision

Revision

Revision

Revision

R

The major version of the Field Diagnostics specification used

for the development of this device.

1

PASSWRD

PASSWRD

PASSWRD

PASSWRD

Password

Password

Password

Password

RW

Password to activate access rights to restricted parameters

0

PROTECT_INFO

PROTECT_INFO

PROTECT_INFO

PROTECT_INFO

access protection

access protection

access protection

access protection

R

Setting:

bit 5...7: not use / User / Service level / Production / Custody

Transfer / Write Lock

Production

SERIAL_NUMBER

SERIAL_NUMBER

SERIAL_NUMBER

SERIAL_NUMBER

Seriennummer

Seriennummer

Seriennummer

Seriennummer

R

Device serial number, serial number of the system.

SW_SIMULATE

SW_SIMULATE

SW_SIMULATE

SW_SIMULATE

SW Simulation

SW Simulation

SW Simulation

SW Simulation

RW

Enable/disable simulation like simulate hardware jumper.

Disabled

Setting:

Disabled / Active

Parameter

Parameter

Parameter

Parameter
DD name

Access

Description and settings

Initial Value