Cleaning, Specifications – Martel Electronics BetaGauge PIR PRO User Manual

Page 10

Advertising
background image

RS-232 Interface

An RS-232 interface is standard on the BetaGauge PIR. Serial
communication can be used for configuration, calibration, and to
transfer measurement data from the gauge. For detailed
specifications on the interface and software communication, see
pages 14 and 15 in this manual.

WARNING

The RS-232 interface must not be used in hazardous areas.

7. Cleaning

To clean the BetaGauge PIR use a cloth with a mild cleaning
solution.

8. Specifications

All specifications cover the temperature range from 0°C to +50°C, unless
otherwise noted.

Available Input Ranges

See page 13 for a table of available ranges in PSI plus equivalent
ranges and resolution for all engineering units

Accuracy

Positive Pressure:: ±0.04% of reading ±0.01% of full scale

Vacuum: ±0.1% of the full scale vacuum measurement

(15 PSI or 1 Bar)

For gauges with full scale ranges equal to, or less than 100 psi
(2 BAR), vacuum operation is limited to -5psi (-350 mBar) except
for the compound range of -15 to 30 PSI (-1 to 2 Bar).

Over Pressure Protection:

See table of Ranges and Resolutions, page 13.

Temperature Compensation

0°C to +50°C (32°F to +122°F) to rated accuracy

Note: For temperatures from –10 °C to 0 °C and 50 °C to 55 °C add
.005% F.S./°C

8